Start with the solvent, not the polymer
We lost a 96-position sequence last March because the septa in a lot of 2 mL vials were silicone, and the autosampler was running 100% ethyl acetate. By injection 40 the peak areas for the internal standard had dropped 18% against the bracketing calibrators, and the GC-MS source was reading siloxane fragments at m/z 73, 147, 207, 281. The vials were fine. The closures were wrong. That is what this guide is about: which polymer belongs in which method, and what happens when you get it wrong....